Errors Without A Message - Dentsply Sirona DAC Universal Operating Instructions Manual

Hide thumbs Also See for DAC Universal:
Table of Contents

Advertisement

Sirona Dental A/S
Operating Instructions DAC Universal
Specifications
Leakage.
There is water in the chamber after end
of cycle.
Condensation in the instruments.
Adapter is loose.
Wrong liquid.
Sterilization OFF safety ring! (A long
acoustic alarm will be given
simultaneously).
The DAC Universal does not open.
The water tank is not automatically
replenished.
A metallic sound is heard when the pump
is activated.
The oil dispenser does not dispense oil. Check whether there is oil in the NitramOil #2 bottle and whether oil is
66 15 574 D3472
D3472.201.20.01.02 03.2019
7.2

Errors without a message

Corrective action
Wipe away any water that may have been spilled after filling.
If the unit continues to leak, please contact your authorized dealer.
Drain the chamber and start a cycle, see "Drain the chamber [ → 88]".
If the error persists, clean or change the drain filter, see "Cleaning the
drain filter [ → 112]" or "Changing the drain filter [ → 120]".
Check whether the drain hose is blocked or damaged.
Empty the drain tank if it is full.
If the error occurs repeatedly, please contact your authorized dealer.
If the instruments are to be stored for an extended period, it is important
to purge them (horizontally) with compressed air to remove condensation
from the instruments. Condensation can lead to rust.
IMPORTANT: Use the compressed air at the pressure level (bar)
recommended by the instrument manufacturer.
The instruments must be wrapped after purging.
Tighten the adapter with the Allen key included.
If a liquid other than demineralized or distilled water has been filled into the
water tank, the DAC Universal must not be used.
Please contact your authorized dealer.
Your DAC Universal technician has run a special program.
Disconnect the plug of the DAC Universal and reconnect it to the power
supply.
WARNING: Otherwise no sterilization will take place!
If the lid holder is stuck on the safety ring, please first reset, see "Reset
[ → 138]".
IMPORTANT: The DAC Universal must not be started without the lid in
the correct position.
Check whether the automatic water intake is activated in the DAC
Universal menu, see "Menu system. [ → 65]".
Check whether water from the external water supply is fed to the DAC
Universal.
If the error occurs repeatedly, please contact your authorized dealer.
Please contact your authorized dealer.
visible in the oil dispenser.
Replace the NitramOil #2 bottle if it is empty, see "Adding/replacing the
lubrication concentrate".
If the error occurs repeatedly, please contact your authorized dealer.
7 Errors / Display messages
7.2 Errors without a message
137