5 Operation
5.5 Main home screen
Pin not correct
Confirm via 'OK' to insert
the pin code again.
"Laser not calibrated" if
the calibration test was not
successful.
Confirmation via 'OK'.
54
5.5.4

Error messages, warnings and instructions

5.5.4.1

Error messages and warnings

The screen is displayed in
case of anomaly. It avoids
hazardous situation result-
ing from failure of the
it-network: downgrade not
allowed or corrupted pack-
age.
Confirmation via 'OK'.
Displayed when the -tem-
perature sensor of the l-
aser module is defec-
tive.Please contact
Dentsply Sirona, your local
dental dealer, or your au-
thorized service center.
Confirmation via 'OK'.
Before battery calibration,
this screen is displayed.
'OK' for confirmation,
'arrow' to go back to the
current screen. No action
occurs.
Displayed when the laser
stop button has been
pressed.
Confirmation via 'OK'.
Dentsply Sirona
Instructions for use SiroLaser Blue
Warns the user that USB
is missing. Please plug a
suitable USB device
(Version 2.0) into the slot.
Confirmation via 'OK'.
65 40 863 D3648
D3648.201.01.07.02    2023-08