List Of Messages; Error Messages; General; Structure - Dentsply Sirona Heliodent Plus Service Manual

Hide thumbs Also See for Heliodent Plus:
Table of Contents

Advertisement

Dentsply Sirona
Service Manual Heliodent Plus
62 15 102 D3507
D3507.076.01.15.02 03.2018
2

List of messages

2.1

Error messages

2.1.1

General

Explanation
The error messages appear on the display of the control electronics.
Recognition
Error messages can be recognized by a six-digit error code (Ex yy zz)
beginning with a large E.
The structure of the error messages is explained in the section entitled
"Structure [ → 25]".
Handling error messages
As a general rule, error messages are acknowledged via all buttons
except for the release button.
If trouble-free operation is possible after the error is acknowledged, then
no further action is necessary.
If error messages reoccur or occur frequently or trouble-free operation is
not possible, identify the error as described in the section "List of error
messages [ → 26]" and take appropriate action to eliminate the
corresponding error or fault.
2.1.2

Structure

Explanation
The codes provide you with error type, error location and troubleshooting
information. Plain text error output follows.
Configuration
The error codes are structured according to the following pattern: Ex yy zz
Explanation of abbreviations:
Ex – Error type
"Remedy" classification for the user. The x character provides a
foundation for making quick decisions as to how serious the error is and
how to handle the error.
yy – Locality
Describes the impacted functionality.
This functionality can be:
● Subassembly
● Subsystem
● Logical functional unit
zz – Identification
Describes a further specification of the error via a consecutive number
with error identification.
2 List of messages
2.1 Error messages
25