Calibration
Calibration Procedure
Output
1. Set the power supply and load operate at full output. You must ensure the power
Current
2. Minimum calibration level
3. Enter current data
4. Maximum calibration level
5. Enter current data
6. Power supply calculates and stores calibration constants.
128
supply is operating in current mode during current calibration. Place a shunt on
the load line so that you can measure the current. Attach a DVM across the shunt
Set the output current to 10% by sending the command:
CAL:OUTP:CURR:LEV MIN
Enter the current read from the external DVM via the shunt.
CAL:OUTP:CURR:DATA <current>
Set the output current to 90% by sending the command:
CAL:OUTP:CURR:LEV MAX
Enter the current read from the shunt via the external DVM.
CAL:OUTP:CURR:DATA <current>
Operating Manual for Multichannel Functionality (GPIB-M)