Read Error Messages; Clear Status; Configure Auxiliary Status Lines - Xantrex GPIB-M-HPD Operating Manual

Multichannel functionality for programmable dc power supplies
Table of Contents

Advertisement

Operation
Power Supply Operation
Read Error
The error queue holds up to 50 error messages. Once each message has been read, it
is cleared from the system.
Messages
To read from the error queue:
SYST:ERR?
This command returns an error code and message from the error queue. For example:
See
Clear Status Issue a Clear Status command to clear all error messages and status registers.
*CLS
or
STATus<channel>:CLEar
Configure
The Auxiliary (Aux) lines are 2 open collector outputs that can be used to monitor
Auxiliary
the status of the power supply. The auxiliary lines are referred to as AUX A and AUX
B. The auxiliary status lines can be set up to report the following status conditions:
Status Lines
58
-315, "Configuration memory test"
-100, "Command error"
0, "No error"
Appendix C. "Error Messages"
NONE
ON: Output On
OFF: Output Off
OVOLtage: Over-Voltage Condition
UVOLtage: Under-Voltage Condition
OCURrent: Over-Current Condition
UCURrent: Under-Current Condition
FOLD: Fold Protection Tripped
CC: Constant Current Mode
CV: Constant Voltage Mode
UNRegulated: Output Unregulated
OTEMperature: Over Temperature Protection Tripper
ACOFf: AC Protection Tripped
SPRotection: Sense Protection Tripper
for descriptions of the messages.