Calibration Procedure

Calibration of programming and readback are combined in a single procedure.
Output
1. Set the load to open circuit. Attach a DVM across the output terminals.
Voltage
2. Minimum calibration level
3. Enter voltage data
4. Maximum calibration level
5. Enter voltage data
6. Power supply calculates and stores calibration constants.
Release 2.1
Set the output voltage to 10% by sending the command:
CAL:OUTP:VOLT:LEV MIN
Enter the voltage read from the external DVM.
CAL:OUTP:VOLT:DATA <voltage>
Set the output voltage to 90% by sending the command:
CAL:OUTP:VOLT:LEV MAX
Enter the voltage read from the external DVM.
CAL:OUTP:VOLT:DATA <voltage>
Calibration
Calibration Procedure
127