8.4
Description
Adjustment step-by-
step
Check & Adjust

V-Index Error

The vertical index does not coincide with the standing axis.
Tilt sensor calibration followed by a crosshairs calibration is recom-
mended before starting V-Index calibration! Refer to "8.2 Tilt Sensor
Calibration" and "8.3 Crosshairs Offset".
Before Calibration:
3D Disto_031
a
Height error
b
Vertical angle offset
Step
Description
1.
Set up the 3D Disto close to a wall with a steep target of good vis-
ibility that is at least 15 m above the instrument.
3D Disto_051
2.
Calibration is started in Menu » Device » Calibration.
3.
Press
.
4.
Pointfinder opens.
Target as precisely as possible.
5.
Press
.
6.
3D Disto turns to second face automatically.
Target a second time. Press
7.
If both measurements were successful, a pop-up window appears
with Set new: XXX gon, Reset to factory default, or Cancel
calibration.
8.
Choose Set... or Reset... to adjust V-index.
After Calibration:
a
b
3D Disto_032
.
79