Appendices
Power supply unit
Reference number
Storage (with and without packaging)
Transport (with and without packaging)
Operation
Input voltage
Mains frequency
13 Appendices
13.1 Symbols Used
Declaration of conformity according to the applicable European directives
Manufacturer
Compliance with the requirements according to "FCC Part 15" (USA)
In some jurisdictions it is not permissible to dispose of these products with unsorted household
waste. Disposal that is not in accordance with the regulations of your country may have a detri
mental impact on health and the environment. Please observe the instructions of your national
authority pertaining to return and collection.
Compliance with the requirements under the "Radiocommunications Act" (AUS)
Serial number
Non-ionising radiation
13.2 Operating states/error signals
13.2.1 Warnings/error signals
Beep signal
1 x long
1 x long
2 x
28
Vibration signal
1 x
5 x
1 x
757L24
-25 °C/-13 °F to +85 °C/+185 °F
10% to 95% relative humidity
-25 °C/-13 °F to +85 °C/+185 °F
10% to 95% relative humidity
-25 °C/-13 °F to +40 °C/+104 °F
Max. 95% relative humidity, non-condensing
90 V~ to 264 V~
47 Hz to 63 Hz
Error
Critical error
(e.g. a sensor is not opera
tional)
Severe error
(e.g. temperature of the lift
ing motor too high)
Malfunction
(e.g.
incompatible
com
ponent connected)
Required action
Turn the product off and on
again or, with the product
turned on, connect the bat
tery
charger.
Contact
Ottobock Service if the
error
has
not
been
resolved.
DynamicArm 12K100N