6.11 Wiring the product with the electric wrist rotator
6.12 Wiring without electric wrist rotator
DynamicArm 12K100N
1) The MyoRotronic cannot be combined with the
elbow joint. The MyoRotronic would be destroyed
when operating the elbow joint.
2) There are numbers for the plug contacts on the
plastic housing of the electric wrist rotator. These
numbers serve as orientation for connecting the
cables.
3) Attach the motor cable to one of the two contacts.
Connect two-pole receptacle to contact no. 3.
4) Connect three-pole receptacle to contact no. 2.
One of the three-pole connectors is marked with a
coloured dot.
5) If the two three-pole connectors are interchanged
when connecting them to the two contacts, no func­
tion of the System Electric Hand or System Electric
Greifer is possible.
6) To secure the coaxial plug, insert the retaining ring
into the groove of the lamination ring using tweez­
ers.
7) Then install the electric wrist rotator in the lamina­
tion ring.
1) One of the two three-pole connectors is marked
with a coloured dot (arrow). The two receptacles
must not be interchanged!
INFORMATION: The two-pole motor cable must
not be attached to the coaxial plug.
Preparation for use