6.2 Gluing in the lamination ring with quick-disconnect wrist unit
1. Lamination cover
2. Lamination ring
3. Socket head screw
4. Dummy for electric wrist rotator
The lamination ring with quick-disconnect wrist unit has to be glued in after the forearm has been cut to length.
Before starting to glue it in, please read these instructions very carefully and proceed exactly in the order
described.
INFORMATION
Read the relevant material safety data sheets before processing the 636K18 Orthocryl sealing resin compact
adhesive and the 617H14 hardener paste.
DynamicArm 12K100N
1) Determine the height of the lamination ring.
2) Mark this measurement on the inside of the fore­
arm.
3) Mix 636K18=1 Orthocryl sealing resin compact
adhesive and 2% 617H14 hardener paste.
4) Cover the outside of the forearm with 627B4 poly­
ethylene adhesive tape to avoid soiling it while glu­
ing.
Preparation for use
11