3D
Duty cycle for height adjustment
Line voltage
Line current
Line harmonics
Max. permissible apparent impedance of supply mains
Max. continuous heat dissipation
Internal fuse(s)
User replaceable
- 1 fuse on permanently
installed X-ray units
- 2 fuses on X-ray units with
detachable power supply cord
Type
External fuse(s)
Battery
Max. weight
2D / 3D s / 3D Classic
3D Plus / 3D Mid
3D Max
Scanning ceph
Planmeca ProCeph
Environmental requirements
User's Manual (3D Imaging)
3D s / 3D Classic: 1.58
3D Plus / 3D Mid / 3D Max:
1.38, 1.40, 1.42, 1.43 or 1.8
25 s ON / 400 s OFF
100 - 220 V~ / 50 - 60 Hz
230 - 240 V~ / 50 Hz
8 - 17 A
Cos better than 0.9
0.5  (100 VAC)
250 W
100 - 220 V~ / 16A FF H 500 V
230 - 240 V~ / 8A FF H 500 V
195100 ELU
100 - 220 V ~ / 16A T 250 V
230 - 240 V ~ / 10A T 250 V
Lithium battery: CR2032, Panasonic
119 kg (262 lb)
141 kg (311 lb)
139 kg (306 lb)
26 kg (57 lb)
20 kg (44 lb)