Planmeca ProMax 3D s User Manual page 61

Hide thumbs Also See for ProMax 3D s:
Table of Contents

Advertisement

(Part number 10030330)
Adjustable
adapter
in highest position
Patient support table
11.1.2 Selecting settings
User's Manual (3D Imaging)
Polystyrene disc
3.
Let the material set. The setting time depends on the
material used. Wait slightly longer than
recommended in the instructions supplied by the
manufacturer to ensure proper hardening.
4.
Remove all excess material from the top of the
calibration cup.
5.
Gently pull the calibration pin out and ensure that the
surfaces of the hole formed by the calibration pin are
even (no air bubbles in inside walls).
6.
Remove any patient supports attached to the X-ray
unit patient support table. Insert the polystyrene disc
provided (part number 10030330) into the adjustable
adapter.
7.
Position the adjustable adapter so that it is in the
highest position.
Refer to section 9.1.3 "Adjusting adapter height" on
page 29 for details.
1.
On the X-ray unit, select the program.
For impression material select
3D Models > Impression.
For plaster material select
3D Models > Plaster Cast.
Planmeca ProMax 3D s & 3D Classic
11 3D MODEL EXPOSURE
57