Code
H151
Line voltage
H152
H161
Temperature
H162
H165
H166
H170
User related
messages
H171
H172
H175
H180
H181
H182
H183
H184
H185
H186
H187
H189
H193
H194
H195
User's Manual (3D Imaging)
Explanation
The line voltage was too low during
exposure.
The line voltage is too low.
The temperature of the tube head is too
high.
The temperature of the lift motor is too
high.
The temperature of the tube head is too
high for the selected exposure values.
The maximum tube head energy level
was exceeded.
Wrong licence code.
The sensor is not attached properly to
the C-arm.
The sensor is not attached properly to
the cephalostat.
PC program selection is in conflict with
the selected X-ray unit program.
DEC is not available.
The imaging process was cancelled in
Planmeca Romexis.
Timeout in image data transmission.
The attached sensor is not suitable for
the selected program.
Remove the 3D sensor.
The 3D sensor is not attached properly. Attach and / or lock the sensor
No IP address defined for 3D sensor.
Problem during image data
transmission.
The screen was touched during
exposure.
Invalid scan settings.
The control panel (touch screen or
ProTouch desktop application) is not
connected to the ProMax main
software.
Request timed out while waiting for
ProMax to respond.
13 HELP MESSAGES
Comments
Exposure was interrupted.
Contact your service technician
for help.
Exposure is not possible.
Contact your service technician
for help.
Wait for a few minutes for the
tube head to cool down.
Wait for a few minutes for the lift
motor to cool down.
Wait for a few minutes for the
tube head to cool down.
Wait for a few minutes for the
tube head to cool down or use
lower exposure values.
Check the licence code.
Attach and / or lock the sensor
in position.
Attach and / or lock the sensor
in position.
Select another exposure mode
in Planmeca Romexis.
Exposure was interrupted.
Contact your service technician
for help.
Change the sensor.
in position.
Exposure was interrupted.
Contact your service technician
for help.
Exposure was interrupted.
Retry or contact your service
technician for help.
Operating in standalone mode.
Retry or contact your service
technician for help.
Planmeca ProMax 3D s & 3D Classic
81