9 3D PATIENT EXPOSURE
PATIENT SIZE
Child (XS)
Small adult (S)
Medium-sized adult (M)
Large adult (L)
Extra large adult (XL)
40
Planmeca ProMax 3D s & 3D Classic
Factory presets for other image resolutions
kV VALUE
If you need to adjust the preset exposure values for this
exposure:
1.
2.
3.
NOTE
You can adjust the preset exposure values permanently as
described in section 12.2.1 "Programs (2100)" on page 75.
mA VALUE
90
5
90
6.3
90
8
90
10
90
12.5
Select the kV / mA field.
Use the minus or plus buttons to set the exposure
values you wish to use. To improve the image
contrast, reduce the kV value. To reduce the
radiation dose, reduce the mA value.
Select the green check mark button.
1
mA VALUE
WITH ULD
3.6
4.5
5.6
7.1
9
2
3
User's Manual (3D Imaging)