9 3D PATIENT EXPOSURE
9.8.2 Moving image volume horizontally (X and Y lasers)
Front light
(X laser)
• Front light (X laser)
Left
Positioning joystick
(front light = X laser)
44
Planmeca ProMax 3D s & 3D Classic
The volume center lights cross in the middle of the image
volume. With the patient positioned in the unit, the volume
center lights form red lines on the front (front light) and on
the left side (side light) of your patient's head.
Check that the image volume is positioned correctly for
your patient. If you need to adjust the volume position,
proceed as follows.
If you need to move the image volume to your left or right:
•
Move the positioning joystick to your left or right. The
front light (i.e. the image volume center as seen from
the front) will move accordingly on your patient's
face.
Right
Left
Side light
(Y laser)
Right
User's Manual (3D Imaging)