Planmeca ProMax 3D s User Manual page 95

Hide thumbs Also See for ProMax 3D s:
Table of Contents

Advertisement

Tube housing front cover
quality equivalent filtration
(not included in the specified total
filtration)
Anode voltage
Pan / SmartPan
Ceph
3D s / 3D Classic /
3D Plus or 3D Mid with Toshiba
D-054SB
3D Max with Toshiba D-067SB
3D Plus / 3D Mid / 3D Max
with Toshiba D-059SBR or
SXR 130-10-0.5 SC
Anode current
Pan
Ceph
3D s / 3D Classic / 3D Plus or
3D Mid with Toshiba D-054SB or
3D Max with Toshiba D-067SB
3D Plus / 3D Mid / 3D Max
with Toshiba D-059SBR
3D Plus / 3D Mid / 3D Max
with Superior SXR 130-10-0.5 SC
mAs range
User's Manual (3D Imaging)
0.3 mm Al @ 70 kV / HVL 2.6 mm Al
60 - 84 kV ±5%
60 - 84 kV ±5%
60 - 90 kV ±5%
60 - 96 kV ±5%
60 - 120 kV ±5%
1 - 16 mA ±10%
1 - 16 mA ±10%
3D
Pan / SmartPan
Scanning ceph
Planmeca ProCeph
3D
Pan / SmartPan
Scanning ceph
Planmeca ProCeph
3D
Pan / SmartPan
Scanning ceph
Planmeca ProCeph
18 TECHNICAL SPECIFICATIONS
1 - 14 mA ±10%
1 - 16 mA ±10%
1 - 16 mA ±10%
16 mA ±10%
1 - 14 mA ±10%
1 - 14 mA ±10%
1 - 14 mA ±10%
14 mA ±10%
1 - 14 mA ±10%
1 - 16 mA ±10%
1 - 16 mA ±10%
16 mA ±10%
Planmeca ProMax 3D s & 3D Classic