Des Comp On/Off - Dräger Fabius Instructions For Use Manual

Mri
Table of Contents

Advertisement

1 During the calibration period, the word "CAL"
replaces the O
value in the oxygen monitoring
2
window. Calibration time is approximately
15 seconds. Upon successful completion of the
calibration, the O
measurement value will be
2
restored.
If the O
sensor cannot be calibrated, replace the
2
O
capsule in the O
sensor housing (see
2
2
page 189).
If the O
sensor still cannot be calibrated, call
2
DrägerService or your local authorized service
organization.

Des Comp On/Off

CAUTION
The Des Comp On/Off function is not relevant for
MRI use.
CAUTION
The Des Comp On/Off parameter remains in the
Off position on the Fabius MRI.
WARNING
The Desflurane vaporizer is not useable in an
MR environment. In an MR environment func-
tionality of the Desflurane vaporizer will be
compromised.
Instructions for Use Fabius MRI SW 3.n
Configuration