Technical Data
MRI Use
Usable within magnets of 1.5 tesla and 3 tesla static magnetic field.
Distance to center of magnetic field: fringe field strength of 40 mtesla (400 gauss) or less.
Ambient Conditions
During operation
Temperature
Atmospheric pressure
Relative humidity
During storage
Temperature
Atmospheric pressure
Relative humidity
The service conditions for supplementary equipment must be noted. These may restrict the area of use
for the system as a whole. Vaporizer units and the anesthetic agents used may restrict the area of use of
the workstation in regards to its temperature range and maximum fresh-gas flow. The corresponding
Instructions for Use of the supplementary equipment must be noted.
Machine Data
Gas supply from medical gas pipeline system
Pipeline System Pressure Range at Machine Connector
O
, N
O, AIR:
2
2
Gas supply connectors:
Each inlet is fitted with a non-return valve
Pipeline Pressure Indicator Accuracy
196
10 to 35 °C (50 to 95 °F)
700 to 1060 cmH
O (hPa)
2
20 to 80 % (no condensation)
–10 to 60 °C (14 to 140 °F)
700 to 1060 cmH
O (hPa)
2
10 to 90 % (no condensation)
50 to 55 psi (3.4 to 3.8 kPa x 100)
Note: Pipeline system supply pressure variation
shall not exceed ±10 %
NIST or DISS (where required)
±3 % of full scale from 40 to 120 psi
(2.7 to 8.3 kPa x 100)
Instructions for Use Fabius MRI SW 3.n