Access Alarm Log

1 Press the »Access Alarm Log« key on the
Standby screen. The Standby soft key window
is replaced by the alarm log, which lists all
alarms with their dates and times.
To scroll through the alarm log, turn the rotary
knob.
To delete all alarms from the log, select and
confirm the »Clear Alarm Log« label.
To exit the alarm log and return to the Standby
screen, select and confirm the return arrow.
CAUTION
Alarm log data is deleted if Fabius MRI is swit-
ched off or a total loss of electrical power supply
occurs.

Restore Site Defaults

2 Press the »Restore Site Defaults« key on the
Standby screen. The predefined site defaults
are restored, and a »Site Default settings
restored« message is displayed above the
Standby soft keys.
The site default settings are password protected.
Changes of the settings can be made via the
Standby Setup screen, see page 140.
WARNING
Risk of inappropriate ventilation settings
After site default settings have been restored
check whether the ventilation and monitoring
settings are appropriate to the patient.
Instructions for Use Fabius MRI SW 3.n
Configuration
1
2