Fault-Cause-Remedy
Priority Alarm Message
!
FLOW SENSOR CAL DUE
(4)
!
FLOW SENSOR FAIL
(8)
!!
FRESH GAS LOW
(21)
!!
INSP O2 HIGH
(13)
!!!
INSP O2 LOW
(31)
!!
INSP PRES NOT REACH
(11)
!!
MINUTE VOLUME HIGH
(14)
164
Probable Cause
More than 18 hours passed
since last flow sensor calibra-
tion.
Cable has been removed and
reconnected.
Sensor cable is disconnected.
Flow sensor has not been
properly calibrated.
Sensor faulty.
Inadequate fresh-gas supply in
all ventilation modes.
Blocked/kinked hose.
Breathing circuit disconnected
or leakage.
Inspiratory O
concentration
2
exceeds the upper alarm limit.
Inspiratory O
concentration is
2
below lower alarm limit.
Plateau pressure is more than
3 cmH
O (hPa) below the
2
P
setting and the expected
INSP
PLAT while ventilating in Pres-
sure Control, Pressure Sup-
port, or SIMV/PS mode.
Minute volume has exceeded
upper alarm limit.
Flow sensor has not been cali-
brated.
Sensor faulty.
Remedy
Follow the procedure to cali-
brate flow sensor (see
page 134).
Reconnect sensor cable to
sensor at breathing system.
Follow the procedure to cali-
brate sensor (see page 134).
Replace sensor and calibrate.
Call DrägerService or your
local authorized service orga-
nization.
Ensure adequate fresh-gas
supply.
Check hoses.
Check breathing circuit.
Check flowmeter settings and
O
high alarm limit.
2
Check O
supply. Check flow-
2
meter settings and O
alarm limit.
Check ventilator, patient cir-
cuit, and P
INSP
Calibrate flow sensor.
Replace if necessary.
Instructions for Use Fabius MRI SW 3.n
low
2
settings.