Dräger Fabius Instructions For Use Manual page 98

Mri
Table of Contents

Advertisement

Operation
If the ventilator settings are correct, confirm the
mode change.
If the ventilator settings are not correct, for each
parameter that needs to change, press the cor-
responding soft key, select the correct value,
and confirm the change. When the parameter
changes are completed, confirm the ventilation
mode change.
After the mode change is confirmed, the
»Volume Control« key LED switches from
blinking to constantly on, the ventilator switches
to the Volume Control mode, and the waveform
is restored.
The parameters that can be set for Volume Control
mode are shown the adjacent table, along with their
adjustment ranges and factory default values.
98
Ventilation Parameter
(Volume Control
Mode)
Pressure Limitation
P
[cmH
O] ([hPa])
MAX
2
Tidal Volume
V
[mL]
T
Frequency
Freq [bpm] ([1/min])
Insp time:
Exp time
T
:T
I
E
Insp pause time:
Insp time
T
:T
[%]
IP
I
PEEP [cmH
O] ([hPa]) 0 to 20
2
Instructions for Use Fabius MRI SW 3.n
Adjustment
Factory
Range
Default
Value
15 to 70
40
min. PEEP+10
20 to 1400
600
4 to 60
12
4:1 to 1:4
1:2
0 to 50
10
0