Reinstalling the Ventilator

1 Open the ventilator door with the attached ven-
tilator unit.
2 Unlatch the three clasps.
3 Remove the cover.
4 Insert the diaphragm.
Fit the cover and lock the three clasps.
5 Connect pressure sensor line of the ventilator
chamber to the respecive connector.
Close the ventilator door with the attached ven-
tilator unit.
180
5
3
5
4
2
1