Gas supply from supplementary O
Cylinder Connections
Cylinder Gas Pressure
(typical full loads at 70 ºF, 21 ºC)
Cylinder Gauges
Cylinder Gauge Range
Compressed gas supply at workstation inlet
Dew point
Oil content
Particles
Internal Regulator Safety Relief Valve Pressure 70 psi (4.8 kPa x 100) opening pressure
Protection Class
Applied parts:
Breathing system (nozzles, breathing hoses)
Ingress of Fluids
Power supply, Rating non-configurable
Rechargeable batteries
Rating:
Type:
Recharging time:
Operation time with fully charged batteries:
Instructions for Use Fabius MRI SW 3.n
and N
O cylinders (with pin-index connections)
2
2
Pin-indexed hanger yokes (CGA V-1-1994)
O
, AIR
2
N
O
2
Conform to ASME B40.1 Grade B
O
2
N
O
2
AIR
>5 °C (41 °F) at ambient temperature
<0.1 mg/m
dust-free air (filtered with pores <1 µm)
I, in accordance with IEC 60601-1
Type BF
IPX0
100 to 240 VAC, 50/60 Hz, 8.7 A including auxiliary
power sockets
24 V; 3.5 Ah
sealed, gelled lead-acid
≤16 hours on the mains or full operation time
45 minutes, minimum
1900 psi (131 kPa x 100)
745 psi (51.3 kPa x 100)
0 to 3000 psi (206.8 kPa x 100)
0 to 3000 psi (206.8 kPa x 100)
0 to 3000 psi (206.8 kPa x 100)
3
Technical Data