Leak/Compliance Test

The »Leak/Compl Test« key is used to initiate a
system compliance test, a system leak test, and a
ventilator leak test.
1 Press the »Leak/Compl Test« key on the
Standby screen. The Standby soft key window
is replaced by a ventilator preparation message
and then instructions for performing the test.
Follow the instructions on the screen:
– Close all fresh gas control valves.
– Place Y-piece on plug located on bag mount.
– Seal sample-line connector (if applicable).
– Switch APL valve to "MAN" and set pressure to
40 to 50.
– Press O
flush button, build up a pressure of
2
15 to 30.
Upon completion of the test, the results are
posted on the screen.
Press the rotary knob to return to the Standby
screen.