Calibrate O 2 Sensor - Dräger Fabius Instructions For Use Manual

Mri
Table of Contents

Advertisement

Configuration
Calibrate O
Sensor
2
To calibrate the oxygen sensor correctly, make sure
it is exposed only to room air during the entire cali-
bration period.
To avoid leakage, remove the oxygen sensor
assembly from the inspiratory valve dome, and
insert the valve dome plug into the dome.
1 Press the »Calibrate O2 Sensor« soft key in
the Setup window.
The calibration instruction window replaces the
Setup window. Follow the directions provided:
– Remove O
Sensor and expose to room air for
2
2 minutes.
– To start O
Calibration press rotary knob.
2
– Observe Calibration status in O
– Reinsert O
Sensor after successful Calibra-
2
tion.
154
data window.
2
1
Instructions for Use Fabius MRI SW 3.n