Volume Alarms On/Off

1 Press the »Volume Alarms On« soft key in the
Setup window. The key label changes from
»Volume Alarms On« to »Volume Alarms
Off«, and the volume alarms are disabled.
NOTE
The »Volume Alarms On/Off« soft key label does
not appear in ManSpont mode because it is select-
able on the ManSpont screen.

Auto Set

2 Press the »Auto Set« soft key in the Setup win-
dow. The breathing pressure threshold is set to
4 cmH
O (hPa) below the current plateau pres-
2
sure data value.
NOTE
The threshold setting may not be less than
5 cmH
O (5 hPa) or greater than 30 cmH
2
(30 hPa).
NOTE
In the absence of a current plateau pressure data
value, pressing the soft key will have no effect.
NOTE
In SIMV/PS mode, the breathing pressure thresh-
old will be set relative to the mandatory ventilation
stroke.
Instructions for Use Fabius MRI SW 3.n
1
O
2
2
Configuration