Assembly
WARNING
For operation in magnetic fields, it is not per-
mitted to connect the vaporizer via hose con-
nectors or tapered connectors with the
anesthesia workstation.
WARNING
Risk of injury
Do not use ferromagnetic keyed filler or drain
adapters or tools when the filling or draining
procedure is carried out in magnetic fields.
Ferromagnetic adapters or tools can be
moved by magnetic attraction.
WARNING
Risk of injury
Vaporizers have to be mounted and removed
outside of the MR room.
Do not exchange any vaporizer in the MR
room.
Vaporizers can be moved by magnetic attrac-
tion.
WARNING
Risk of injury
Do not use filling adapters without a label
"MRI". Adapters can be moved by magnetic
attraction.
CAUTION
Only Vapor 2000 vaporizers can by used on the
Fabius MRI in MRT scanner rooms.
CAUTION
Risk of injury
Dräger metal keyed filler adapters and Dräger fill
adapters are labeled with "MRI" are not ferromag-
netic.
73
Instructions for Use Fabius MRI SW 3.n