Dalsa Linea SWIR GigE Series User Manual page 54

Ingaas line scan camera
Table of Contents

Advertisement

Display Name
Defective Pixel
Replacement Mode
Off
Active
Defective Pixel
Replacement Map
Current Active Set
Factory Map
User Map 1
Defective Pixel
Replacement Algorithm
Method 1: Average Two
Adjacent Pixels
Method 3: Neighboring
Pixel
Flat Field Correction
Mode
Off
Active
Calibration
Flat Field Correction
Current Active Set
Factory Flatfield
User Flatfield 1
User Flatfield 2
User Flatfield 3
User Flatfield 4
Flat Field Correction
Type
Line-Based
Flat Field Correction
Algorithm
Method 1
Calibration Algorithm
Basic
52
Operational Reference
Feature & Values
defectivePixelReplacementMode
defectivePixelReplacementMapCurren
tActiveSet
FactoryMap
UserMap1
defectivePixelReplacementAlgorithm
Method1
Method3
flatfieldCorrectionMode
Calibration
flatfieldCorrectionCurrentActiveSet
FactoryFlatfield
UserFlatfield1
UserFlatfield2
UserFlatfield3
UserFlatfield4
flatfieldCorrectionType
LineBase
flatfieldCorrectionAlgorithm
Method1
flatfieldCorrectionCalibrationAlgorith
m
Description
Sets the mode for the defective pixel
replacement.
Off
Defective Pixel Replacement is disabled.
Active
Defective Pixel Replacement is enabled.
Sets the defective pixel replacement set.
Sets the factory coefficient table as active.
Sets the User Map coefficient table as active.
Specifies the defective pixel replacement
algorithm.
This algorithm replaces a defective pixel with the
average value of the pixel to the left and right of
the pixel to be replaced.
This algorithm replaces a defective pixel with a
neighbor.
Sets the mode for the Flat Field correction.
Off
Flat Field Correction is disabled.
Active
Flat Field Correction is enabled.
When selected, the camera is configured for flat
field correction calibration. The device may
automatically adjust some of its features when
calibrate mode is enabled. The features that are
automatically adjusted are device specific. The
device will not restore these features when the
Flat Field Correction Mode feature is changed
from Calibrate mode to another mode.
Specifies the current set of Flat Field coefficients
to use.
Sets the factory Flat Field coefficient table as the
current Flat Field.
Sets User Flat Field 1 coefficient table as the
current Flat Field.
Sets User Flat Field 2 coefficient table as the
current Flat Field.
Sets User Flat Field 3 coefficient table as the
current Flat Field.
Sets User Flat Field 4 coefficient table as the
current Flat Field.
Specifies the Flat Field correction type.
Flat Feld correction is based on an individual line
(FlatLine).
Specifies the Flat Field correction algorithm to
use.
The following formula is used to calculate the
Flat Field corrected pixel: newPixelValue[x] =
(sensorPixelValue[x] - FFCOffset[x]) *
FFCGain[x]
Specifies the Flat Field calibration algorithm to
use.
Basic
Direct calculation of coefficients based on
average line values and target.
Device Version
& View
Beginner
Beginner
Linea SWIR GigE Series Camera
1.00
Expert
DFNC
1.00
Expert
DFNC
1.00
Expert
DFNC
1.00
DFNC
1.00
DFNC
1.00
Guru
DFNC
1.00
Guru
DFNC
1.00
Guru
DFNC