Internal Programmable Exposure
The Linea SWIR GigE camera in the Internal Programmable Exposure mode has the following
features:
The Trigger Source feature (see I/O Control category) selects an internal signal as trigger.
The programmable internal trigger maximum line rate limit is related to the ExposureTime
feature.
Exposure duration is user programmable (exposure maximum is dependent on the line rate).
Minimum exposure (in µs) is model dependent.
External Programmable Exposure
The External Programmable Exposure mode is similar to Internal Programmable except the
exposure start being an external user input.
The Trigger Source feature (see I/O Control category) selects an external signal line as trigger.
Line rate and exposure limits are defined by Internal Programmable Exposure.
Exposure Delay Constraints
Exposure maximum delay specified by the user should not exceed the period of the signal.
If the exposure trigger is edge rather than level, the delay should not exceed half the period.
If these parameters are violated, the camera ignores new transitions / pulses / triggers until
the specified delay has passed.