4
Measurement Operations & Settings
Response:
None
Example:
:trig0:inp ign
Affects:
All power meters
Command:
:TRIGger[n]:INPut:EDGe?
Syntax:
:TRIGger[n]:INPut:EDGe?
Description:
Returns the trigger edge detection of the BNC TTL input.
Parameters:
None
Response:
RISing: trigger on the high edge of the BNC TTL input
FALLing: trigger on the low edge of the BNC TTL input
Example:
:TRIG:INP:EDG? → RIS
Affects:
All power meters
Command:
:TRIGger[n][:INPut:EDGe
Syntax:
:TRIGger[n][:INPut:EDGe
Description:
Sets the trigger edge detection of the BNC TTL input.
Parameters:
0 or RISing: trigger on the high edge of the BNC TTL input
1 or FALLing: trigger on the low edge of the BNC TTL input
Response:
None
Example:
:TRIGg:INP:EDG RIS
Affects:
All power meters
Command:
:TRIGger[n]:OFFSet?
Syntax:
:TRIGger[n]:OFFSet?
Description:
Returns the number of incoming triggers received before data logging begins.
Parameters:
None