Keysight N774 C Series Programming Manual page 73

Multiport power meter
Table of Contents

Advertisement

Response:
A hexadezimal integer value which represents the result for all 4 power meter channels of the selected quad.
Each hexadecimal digit represents one channel .
0: zero succeeded without errors.
Any other number: remote zeroing failed
example 1: sens1:chan:corr:coll:zero:quad? → +272
272 decimal = 0x110. This means zeroing failed on channels 2 and 3. All other channels were successful.
example 2: sens5:chan:corr:coll:zero:quad? → +272
272 decimal = 0x110. This means zeroing failed on channels 6 and 7. All other channels were successful.
Example:
:sens1:chan:corr:coll:zero:quad? → 0
Affects
All power meters
Command:
:SENSe[n]:FUNCtion:LOOP
Syntax:
:SENSe[n]:FUNCtion:LOOP<wsp><value>
Description:
Sets the number of logging loops
Parameters:
Number of Loops, an integer value.
0 = Endless Streaming
1 = 1 (Default)
2 = 2 (For 2 Million Points with Buffer A and B) ....
n
Response:
None
Example:
:SENS1:FUNC:LOOP 0
Affects
All power meters
Command:
:SENSe[n]:FUNCtion:LOOP?
Syntax:
:SENSe[n]:FUNCtion:LOOP?
Description:
Gets the number of logging loops.
For details look at description and example in Using data buffers for simultaneous measurement and upload.
Parameters:
None
Keysight N774xC Series Multiport Power Meter Programming Guide
Measurement Operations & Settings
4
73

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N7744cN7745c

Table of Contents