Keysight N774 C Series Programming Manual page 83

Multiport power meter
Table of Contents

Advertisement

Response:
The averaging time as a float value in seconds.
Example:
:sens1:pow:atim? → +1.00000000E+000
Affects
All power meters
Command:
:SENSe[n]:POWer:RANGe:AUTO
Syntax:
:SENSe[n]:POWer:RANGe:AUTO <wsp><boolean>
Description:
Enables or disables automatic power ranging for the channel.
If automatic power ranging is enabled, ranging is automatically determined by the instrument. Otherwise, it must be set by
the sens[n]:pow:rang command.
Automatic ranging while other commands are sent to power meters has lead to timing conflicts in some configurations.
Automation programs can often better control the range directly.
Parameters:
A boolean value:
Response:
None
Example:
:sens1:pow:rang:auto 1
Affects
All power meters
Command:
:SENSe[n]:POWer:RANGe:AUTO?
Syntax:
:SENSe[n]:POWer:RANGe:AUTO?
Description:
Returns whether automatic power ranging is being used by the channel.
Parameters:
None
Response:
A boolean value:
Example:
:sens1:pow:rang:auto? → 1
Affects
All power meters
Keysight N774xC Series Multiport Power Meter Programming Guide
0 or OFF: automatic ranging disabled
1 or ON: automatic ranging enabled
0 or OFF: automatic ranging disabled
1 or ON: automatic ranging enabled