Keysight N774 C Series Programming Manual page 77

Multiport power meter
Table of Contents

Advertisement

Parameters:
Total time: The total time from the start of stability mode until it is completed.
Period time: A new measurement is started after the completion of every period time.
Averaging time: A measurement is averaged over the averaging time.
NOTE: Setting parameters for the Stability function sets some parameters, including hidden parameters, for the MinMax and
Logging functions and vice versa.
The total time should be longer than the period time.
The period time should be longer than the averaging time.
The number of data points is equal to the total time divided by the period time.
Total time, period time, and averaging time are time values in seconds.
If you specify no units in your command, seconds are used as the default.
See :SENSe[n]:FUNCtion:PARameter:STABility? on page 100 for information on starting/stopping a data acquisition function.
See :SENSe[n]:FUNCtion:RESult? on page 100 for information on accessing the results of a data acquisition function.
Response:
None
Example:
:sens1:func:par:stab 1s,0.1s,0.1s
Affects
All power meters
Command:
:SENSe[n]:FUNCtion:PARameter:STABility?
Syntax:
:SENSe[n]:FUNCtion:PARameter:STABility?
Description:
Returns the total time, period time, and averaging time for the stability data acquisition function.
Parameters:
None
Response:
Total time, delay time, and averaging time are float values in seconds.
Example:
:sens1:func:par:stab? → +1.00000000E+000, +1.00000000E-001,+1.00000000E-001
Affects
All power meters
Command:
:SENSe[n]:FUNCtion:RESult?
Syntax:
:SENSe[n]:FUNCtion:RESult?
Description:
Returns the data array of the last data acquisition function.
Parameters:
None
Keysight N774xC Series Multiport Power Meter Programming Guide
Measurement Operations & Settings
4
77