Keysight N774 C Series Programming Manual page 82

Multiport power meter
Table of Contents

Advertisement

4
Measurement Operations & Settings
Response:
None
Example:
:sens1:pow:gain:auto 1
Affects
All power meters
Command:
:SENSe[n]:POWer:GAIN:AUTO?
Syntax:
:SENSe[n]:POWer:GAIN:AUTO?
Description:
Get the Auto Gain.
Parameters:
None
Response:
0 = Auto Gain Off.
This is the position for best transient response.
1 = Auto Gain On (Default)
This is the Position for best dynamic.
Example:
:sens1:pow:gain:auto? → 1
Affects
All power meters
Command:
:SENSe[n]:POWer:ATIMe
Syntax:
:SENSe[n]:POWer:ATIMe<wsp><averaging time>[NS|US|MS|S]
Description:
Sets the averaging time.
Parameters:
The averaging time as a float value in seconds.
If you specify no units in your command, seconds are used as the default.
NONE: For N774xC power meters the internal granularity of the averaging time is 1us which allows flexibility.
Response:
None
Example:
:sens1:pow:atim 1s
Affects
All power meters
Command:
:SENSe[n]:POWer:ATIMe?
Syntax:
:SENSe[n]:POWer:ATIMe?
Description:
Returns the averaging time.
Parameters:
None