Keysight N774 C Series Programming Manual page 75

Multiport power meter
Table of Contents

Advertisement

Parameters:
Data Points:
Data Points is the number of samples that are recorded before the logging mode is completed.
Data Points is an integer value.
Averaging time:
Averaging time is a time value in seconds. There is no time delay between averaging time periods.
Use :SENSe[n]:FUNCtion:PARameter:STABility? if you want to use delayed measurement.
NOTE: Setting parameters for the logging function sets some parameters, including hidden parameters, for the stability and
MinMax functions and vice versa.
If you specify no units for the averaging time value in your command, seconds are used as the default.
See :SENSe[n]:FUNCtion:STATe for information on starting/stopping a data acquisition function.
See :SENSe[n]:FUNCtion:RESult? for information on accessing the results of a data acquisition function.
NOTE: Before using this command, ensure to stop logging for all available channel.
Details can be found in the Application Note "Transient Optical Power Measurements with the N7744A and N7745A"
http://literature.cdn.keysight.com/litweb/pdf/5990-3710EN.pdf.
Response:
None
Example:
:sens1:func:par:logg 64,1ms
Affects
All power meters
Command:
:SENSe[n]:FUNCtion:PARameter:LOGGing?
Syntax:
:SENSe[n]:FUNCtion:PARameter:LOGGing?
Description:
Returns the number of data points and the averaging time for the logging data acquisition function.
Parameters:
None
Response:
Returns the number of data points as an integer value and the averaging time, t
Example:
:sens1:func:par:logg? → +64,+1.00000000E-001
Affects
All power meters
Command:
:SENSe[n]:FUNCtion:PARameter:MINMax
Syntax:
:SENSe[n]:FUNCtion:PARameter:MINMax<wsp> CONTinous|WINDow|REFResh,<data points>
Description:
Sets the MinMax mode and the number of data points for the MinMax data acquisition function.
Keysight N774xC Series Multiport Power Meter Programming Guide
Measurement Operations & Settings
, as a float value in seconds.
avg
4
75