Keysight N774xC Series Multiport Power Meter Programming Guide
Error
Number
String
-185
"Subop out of range"
Description:
Suboperations are parameters that are passed to refine the destination of a command. They are
used to address slots, channels, laser selections and GPIB/SCPI register levels. This error is
generated if the parameter is not valid in the current context or system configuration.
Example:
This error occurs if the user queries the status of a summary register and passes an invalid status
level.
Note:
Incorrect slots and channels addresses are handled by error code -301
-200
"Execution error (StatExecError)"
Description:
This error occurs when the current function, instrument or module state (or status) prevents the
execution of a command. This is a generic error which can occur for a number of reasons.
Example:
When a powermeter has finished a logging application and data is available, the user is not able to
reconfigure the logging application parameters. First, the user must stop the logging application.
-201
"Invalid while in local"
"Please be patient - GPIB currently locked out"
Description:
Some operations block the complete system. Since no sensible measurements are possible while
this is true, the GPIB is locked out.
Example:
When ARA, Lambda zeroing or zeroing is executing on a TLS module, the GPIB is not accessible.
-202
"Settings lost due to rtl"
-203
"Command protected"
-210
"Trigger error"
-211
"Trigger ignored"
Description:
A trigger has been detected but ignored because of timing contraints. (For Example: average time
to large).
-212
"Arm ignored"
Description:
The user can set the automatic re-arming option for input and output trigger events (see
Strings
on page 96). When this error occurs, the device ignores the setting because the current
module status does not allow the change of trigger settings.
-213
"Init ignored"
Description:
The INIT:IMM command initiates a trigger and completes a full measurement cycle. The continuous
measurement must be DISABLED. This error code is generated if the module is still in cont.
measurement mode.