Parameters:
TOMODule:
TOREF:
The reference as a float value.
You must append a unit type
•
•
The two reference values are completely independent. When you change the reference mode using the command
:SENSe[n]:POWer:REFerence:STATe:RATio, the instrument uses the last reference value entered for the selected reference
mode.
Response:
None
Example:
:sens1:pow:ref tomod,-40DB
Affects
All power meters
Command:
:SENSe[n]:POWer:REFerence?
Syntax:
:SENSe[n]:POWer:REFerence?<wsp>TOMODule|TOREF
Description:
Returns the channels reference value.
Parameters:
TOMODule:
TOREF:
Response:
The reference as a float value.
Example:
:sens1:pow:ref? toref → +1.00000000E-006
Affects
All power meters
Command:
:SENSe[n]:POWer:REFerence:DISPlay
Syntax:
:SENSe[n]:POWer:REFerence:DISPlay
Description:
Takes the current measured input power level value as the reference value.
"disp to ref"
Parameters:
None
Keysight N774xC Series Multiport Power Meter Programming Guide
dB if you use TOMODule or
Watts or dBm if you use TOREF.
Sets the reference value in dB used if you choose measurement relative to another
channel
Sets the reference value in Watts or dBm if you choose measurement relative to a
constant reference value
Returns the reference value in dB used if you choose measurement relative to another
channel
Returns the reference value in Watts or dBm if you choose measurement relative to a
constant reference value
Measurement Operations & Settings
4