Measurement Operations & Settings
Parameters:
CONTinous: continuous MinMax mode
WINDow: window MinMax mode
REFResh: refresh MinMax mode
NOTE: WINDow mode has the same function as REFResh mode. It is included to ensure compatibility.
NOTE: The time between samples in MinMax mode is at least the averaging time, but not less than about 2 ms, depending on
the interface command rate.
Data Points is the number of samples that are recorded in the memory buffer used by the WINDow and REFResh modes. Data
Points is an integer value.
NOTE: Setting parameters for the MinMax function sets some parameters, including hidden parameters, for the stability and
Logging functions and vice versa.
See :SENSe[n]:FUNCtion:STATe for information on starting/stopping a data acquisition function.
See :SENSe[n]:FUNCtion:RESult? for information on accessing the results of a data acquisition function.
See Trigger Subsystem Commands for information on how triggering affects data acquisition functions.
Response:
None
Example:
:sens1:func:par:logg 64,1ms
Affects
All power meters
Command:
:SENSe[n]:FUNCtion:PARameter:MINMax?
Syntax:
:SENSe[n]:FUNCtion:PARameter:MINMax?
Description:
Returns the MinMax mode and the number of data points for the MinMax data acquisition function.
Parameters:
None
Response:
CONT: continuous MinMax mode
WIND: window MinMax mode
REFR: refresh MinMax mode
NOTE: WINDow mode has the same function as REFResh mode. It is included to ensure compatibility. The number of data
points is returned as an integer value.
Example:
:sens1:func:par:minm? → WIND,+10
Affects
All power meters
Command:
:SENSe[n]:FUNCtion:PARameter:STABility
Syntax:
:SENSe[n]:FUNCtion:PARameter:STABility<wsp> <total time>[NS|US|MS|S],<period time>[NS|US|MS|S],<averaging
time>[NS|US|MS|S]
Description:
Sets the total time, period time, and averaging time for the stability data acquisition function.