System Communicate - The :Syst:communicate Sub Tree; System:communicate:ethernet:autoip:enable - Keysight N774 C Series Programming Manual

Multiport power meter
Table of Contents

Advertisement

3
Instrument Setup and Status

System Communicate - The :SYST:COMMunicate sub tree

NOTE
Command:

:SYSTem:COMMunicate:ETHernet:AUTOip:ENABle?

Syntax:
:SYSTem:COMMunicate:ETHernet:AUTOip:ENABle?
Description:
Check whether Automatic IP addressing is enabled or disabled.
Parameters:
None
Response:
Boolean (0 | 1)
Example:
:SYST:COMM:ETH:AUTO:ENAB? -> 1
46
We recommend you change network settings using the local user
interface.
The instrument does not close open connections when restarting the network
interface (:SYSTem:COMMunicate:ETHernet:RESTart). This means the number of
possible connections is reduced by the number of previously open connections.
However, the instrument does make sure connections are still alive. It should
release unused open connections after about two minutes.
Some notes on DHCP/AutoIP/DNS
If DHCP is enabled but no DHCP server is found, the instrument tries to
use AutoIP as a fallback. This may take about 2 minutes.
Depending on the available network capabilities, the instrument tries to
tell the DNS server its host name or read the host and domain named it
has been assigned.
MAC address:
The Media Access Control (MAC) number is a unique number associated
with each network adapter.