4
Measurement Operations & Settings
Parameters:
None
Response:
A hexadezimal integer value which represents the result for all channels.
Each hexadecimal digit represents one channel .
0: zero succeeded without errors.
Any other number: remote zeroing failed
example 1: sens:chan:corr:coll:zero:all? → +272
272 decimal = 0x110. This means zeroing failed on channels 2 and 3. All other
channels were successful.
example 2: sens:chan:corr:coll:zero:all? → +286326784
286326784 decimal = 0x11110000. This means zeroing failed on channels 5,6,7,8.
All other channels were successful.
Example:
:sens:chan:corr:coll:zero:all? → 0
Affects
All power meters
Command:
:SENSe[n]:CORRection:COLLect:ZERO:QUAD
Syntax:
:SENSe[n]:CORRection:COLLect:ZERO:QUAD
Description:
Zeros the electrical offsets for all 4 power meter channels of the selected quad
if [n] is between 1 and 4 then the first quad is zeroed
if [n] is between 5 and 8 then the second quad is zeroed
NOTE: Cover the Optical Inputs or switch the input source off before starting.
Parameters:
None
Response:
None
Example:
:sens1:corr:coll:zero:quad
Affects
All power meters
Command:
:SENSe[n]:CORRection:COLLect:ZERO:QUAD?
Syntax:
:SENSe[n]:CORRection:COLLect:ZERO:QUAD?
Description:
Returns the status of the most recent zero command for all 4 power meter
channels of the selected quad.
if [n] is between 1 and 4 then the first quad zero result is returned
if [n] is between 5 and 8 then the second quad zero result is returned
The result is backed up in the nonvolatile RAM.
NOTE: If a channel fails to zero, it continues to use the result of the last successful zeroing.
Parameters:
None
72
Keysight N774xC Series Multiport Power Meter Programming Guide