Disk Drive Removal Procedure

Instruments with Non-Removable Drives

Instruments with Non-Removable Drives
To remove the disk drive, follow the procedure below, while referring to
7-1. The numbered items in the figure correspond to the step numbers in the
procedure.
Before commencing this procedure, remove the complete Processor Assembly
from the instrument, following the procedure described in the chapter
Processor Assembly Removal
Figure 7-1
Disk Drive Removal Procedure
1. Remove 15 screws from the top lid of the Processor Assembly.
2. Tilt the lid up.
3. While holding the disk drive to prevent it from falling, remove the 4 drive retaining
Security Features & Document of Volatility
screws.
Procedure.