Replacement of Disk Drive (Instruments with PC6S/PC7S)

IMPORTANT
Ensure that the Instrument Software Revision on the secondary
non-classified disk drive matches that of the classified drive in the controlled
area.
Security Features & Document of Volatility
Task
Install secondary non-classified Disk Drive #2, which was previously prepared in
Step 12.
Power on instrument.
When the instrument has booted up, press System, Show System, and confirm
that the Instrument Software Revision is the same as noted in Step 16. If not, upgrade
the instrument software to the same version as recorded in Step 16.
Instrument is now operational with original calibration data. Deliver to service center
for cal/repair.
Service center may or may not generate new calibration files on Disk Drive, depending
on whether an adjustment is performed.
This procedure assumes that new cal data was generated.
Instrument returned to customer's non-controlled area.
Verify Instrument Software Revision, to determine whether the service center upgraded
the software. See Step 2.
Back up (archive) calibration file from Disk Drive to portable read-only media.
For details of the procedure, see
Files" on page
42.
Remove the non-classified Disk Drive #2 and retain outside of controlled area. If
necessary, update the documented software revision on the drive label.
For details of how to remove the Disk Drive, see
page
52.
Physically deploy instrument back into controlled area, without a Disk Drive.
Bring read-only media containing calibration file (created in Step 25) into controlled
area.
Re-insert classified Disk Drive #1 into instrument, and verify Instrument Software
Revision, as in Step 2.
If the Instrument Software Revision does not match the revision documented in step
24
(because the software was updated at service center), then a software update must
be performed, to force the controlled-area drive to match the non-controlled-area
drive that was used to calibrate the instrument.
Restore calibration file (Step 27) from read-only media to instrument.
Done. Go to Step 15.
"Archiving and Restoring Factory Calibration Data
"Disk Drive Removal Procedure" on
37