Advertisement

Memory Sanitization and Removal Procedures
Instrument Sanitization Procedures
If the non–classified drive receives a software upgrade, or if the non-classified
drive is replaced with a drive that contains a newer revision of software than
that of the classified drive, the classified drive will require a software upgrade
(inside the controlled area) to match the non-classified drive.

Items Required

To perform these tasks in a controlled area, you will need:
— 1x Instrument
— 2x Disk Drives for the instrument.

Procedure

Steps below that appear on a yellow background take place in the controlled
area. All other steps take place in a non-controlled area.
N O T E
The first twelve steps listed below are performed in order to prepare a spare
drive (Disk Drive #2) that has the same software revision and calibration files,
and has been used in the same CPU assembly, as the original disk drive (Disk
Drive #1).
Step
1
2
3
4
Security Features & Document of Volatility
One drive (Disk Drive #1) stays permanently within the controlled area, and
the other (Disk Drive #2) stays permanently outside it
Task
Receive new instrument from factory.
The instrument shipment may include a spare disk drive (additional drive) prepared by
the factory, which contains the same calibration data and software revision as the disk
drive in the instrument. If such a factory-prepared disk drive is present, you can skip
steps 3, 4, 6, 7, 9, and
Power up the instrument and verify the Instrument Software Revision.
For instructions on how to verify the currently installed software revision, see
"Determining Instrument Software Revision" on page
Write down the software revision.
If the instrument's software is not the latest version, update the instrument software to
the latest version available. For instructions, see
up-to-date?" on page 3
Write down the software version.
If software update was performed in step 3, back up (archive) the instrument factory
calibration data to internal Flash memory.
For details of the procedure, see
Wizard" on page
47.
10
below.
"Is your product software
"Data Backup or Restore using Alignment Data
43.
38