Overview Of Adjustment Parameters In Mymodes - Otto Bock C-Leg 3C98-3 Instructions For Use Manual

Hide thumbs Also See for C-Leg 3C98-3:
Table of Contents

Advertisement

Use
The following parameters can be modified:
Parameter
Resistance
1
Stance function
1
Sitting function
Acoustic feedback signal
Volume
1
To use these functions in the Cockpit app, they need to be enabled in the adjustment software.

9.2.3 Overview of adjustment parameters in MyModes

The parameters in the MyModes describe the static behaviour of the prosthesis for a specific motion pattern such
as cross-country skiing. Damping behaviour is not automatically controlled and adjusted in MyModes.
The following parameters can be modified in MyModes:
Parameter
Basic flex.
Gain
Basic ext.
Locking angle
Volume
40
Adjustment soft­
Setting range,
ware range
app
120 to 190
+/- 10 of the con­
figured value
0/Off – deactiva­
ted
1/On – activated
0/Off – deactiva­
ted
1/On – activated
On/Off
0 to 4
0 to 4
Adjustment soft­
Setting range,
ware range
app
0–200
+/- 20 of the con­
figured value
0–100
+/- 10 of the con­
figured value
0–60
+/- 20 of the con­
figured value
0–90
+/- 10 of the con­
figured value
0–4
0–4
Meaning
Flexion resistance while sitting down, in the
stance phase, while walking on ramps and
stairs.
Information about this function is provided in
the section "Stance function" (see page 35)
When the function is activated, the resistance
in the flexion direction while sitting is reduced
in addition to the reduction of resistance in
the extension direction.
Acoustic feedback for switching between the
stance and swing phase.
Volume of beep signal for confirmation tones
(e.g. when checking the charge level, switch­
ing MyModes). The "0" setting deactivates
the audible feedback signals. However,
warning signals are still generated if errors
occur.
Meaning
Level of flexion resistance when the knee joint
begins to flex
Increase in flexion resistance (starting with
the "Basic flex." parameter) when flexing the
knee joint. The knee joint locks at a certain
flexion angle, which depends on the settings
for the "Basic flex." and "Gain" parameters.
Level of extension resistance
Angle up to which the knee joint can be
extended.
Information: If this parameter is >0, the knee
joint is locked in a flexed position in the exten­
sion direction. To unlock it, take all weight off
the prosthesis and tilt it back for at least
2 seconds. This enables extension of the joint
independently of the settings for the "Basic
ext." and "Locking angle" parameters. This
may be necessary to switch to basic mode
using a movement pattern.
Volume of beep signal for confirmation tones
(e.g. when checking the charge level, switch­
ing MyModes). The "0" setting deactivates
the audible feedback signals. However,
warning signals are still generated if errors
occur.
C-Leg 3C98-3/3C88-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

C-leg 3c88-3

Table of Contents