Assembly and preparation
If installing the BiliLux with optional spring arm on a Caleo incubator or a Babytherm
radiant warmer, attach it to the 38 mm pole (see "Attaching the BiliLux to a 38 mm
pole" on page 32). Do not attach to any of the DIN or Fairfield rails.
Do not install the BiliLux phototherapy light directly on the Babytherm canopy.
6.3
Placing the BiliLux phototherapy light on an incubator
hood
Position the phototherapy light on the incubator hood with the long side of the light
parallel to the long side of the hood. The four corners of the light have slip-resistant
feet on the bottom that help to keep the light in place.
6.4

Installing the BiliLux with optional spring arm

The BiliLux with spring arm consists of the phototherapy light, the spring arm, and
mounting adapters for a 38 mm pole or for GCX, DIN, and Fairfield rails.
6.4.1
Attaching the BiliLux to a 38 mm pole
1 Attach the compact rail (A) or the clamp (B) to the 38 mm pole.
2 Attach the rail clamp for the phototherapy light (C) to the compact rail (A) or to
the clamp (B).
3 Insert the post (D) on the bottom of the spring arm into the rail clamp (C). Ensure
that the arm is seated properly, and then tighten the adjustment screw (E) on the
rail clamp to secure the arm.
4 Insert the quick-connect plug (F) on the phototherapy light into the arm socket
(G) until it audibly clicks into place.
32
Instructions for use BiliLux