Using Phototherapy (Optional) - Dräger Babytherm 8000 WB Instructions For Use Manual

Hide thumbs Also See for Babytherm 8000 WB:
Table of Contents

Advertisement

Operation

Using phototherapy (optional)

WARNING
Only use the phototherapy lamps indicated, see
"Technical Data", page 37. Using lights of other
types can affect the safety and effectiveness
of the phototherapy.
WARNING
Always use patient eye protection when using
phototherapy.
Keep a constant check that the eye protection
is correctly positioned.
WARNING
Always monitor the differential diagnosis for the neonate
to ensure that no life-saving measure is delayed.
WARNING
The patient's bilirubin levels must be measured regularly.
Beware of possible toxic effects due to bilirubin
photoisomers.
WARNING
Phototherapy can upset the patient's water balance.
WARNING
Dangerous body temperatures can occur when using
reflective foil.
CAUTION
Direct eye contact with the radiation source for an extended
period when in use should be avoided.
CAUTION
Protect patients in the immediate vicinity of the phototherapy
device from the radiation emitted, e.g. by using goggles
and shields.
CAUTION
Users should avoid remaining for an extended time
in the area of the radiation. Risk of sunburn.
NOTE
Use of the bed canopy in conjunction with operation of
the integrated phototherapy system leads to a significant
reduction in the bilirubin-effective radiant intensity and
a very uneven distribution of the radiation on the bed.
28
In manual mode, reduce the heat output of the radiant warmer.
The phototherapy light directs additional warmth to the patient.
CAUTION
Monitor the patient's core temperature regularly using
a separate thermometer.
Swing the radiant warmer over the patient. Position
the phototherapy lights vertically above the patient, as
otherwise the effect of the phototherapy will be diminished.
NOTE
Follow the instructions for use of the phototherapy unit.