Dräger BiliLux Instructions For Use Manual page 44

Led phototherapy light
Hide thumbs Also See for BiliLux:
Table of Contents

Advertisement

Operation
8.1.2
Positioning the BiliLux with spring arm and BiliLux with trolley
at an incubator
1 If the device is equipped with a trolley, adjust the height of the trolley as desired
using the adjustment knob. If equipped with spring arm without trolley, proceed
to step 4.
2 Roll the trolley under the incubator. Keep the distance between the trolley and
the incubator to a minimum.
3 Lock the castors to prevent movement of the trolley.
CAUTION
Risk of injury
Cables or attached accessories may be inadvertently disconnected or otherwise
compromised when adjusting the spring arm.
► Take care not to compromise cables or accessories when adjusting the spring
arm to the desired position.
4 Adjust the spring arm to the desired height and position.
CAUTION
Risk of injury
The LED panel on the phototherapy light may be hot to the touch.
► Adjust the phototherapy light only by touching the housing. Do not touch the
LED panel.
5 Position the phototherapy light over the incubator hood. Ensure that as much of
the patient's skin as possible is covered with phototherapy light.
44
≥30 cm
(12 in)
≥30 cm
(12 in)
Instructions for use BiliLux