11.4
Reprocessing list
Components
Surface
disinfection
with cleaning
Phototherapy
Yes
light
Spring arm
Yes
Trolley
Yes
Radiometer
Yes
11.5
Reprocessing procedures
11.5.1
Validated reprocessing procedures
At the time of product-specific validation, the following reprocessing procedures
showed good material compatibility and effectiveness:
Procedure
Agent
Surface
Dismozon
disinfection with
plus
cleaning
The effectiveness of the listed reprocessing procedures has been validated by
independent laboratories that are certified to the standard ISO 17025.
11.5.2
Disinfectants
Use disinfectants that are nationally approved and are suitable for the particular
reprocessing procedure.
11.5.2.1
Surface disinfectants
At the time of the test, the surface disinfectants listed in the following table showed
good material compatibility. They can be used in addition to the surface
disinfectants listed in the section "Validated reprocessing procedures".
The manufacturers of the surface disinfectants have verified at least the following
spectra of activity:
– Bactericidal
– Yeasticidal
– Virucidal or virucidal against enveloped viruses
Observe the specifications of the surface disinfectant manufacturers.
Other surface disinfectants are used at one's own risk.
Instructions for use BiliLux
Manual
Machine
cleaning
cleaning with
followed by
thermal
disinfection by
disinfection
immersion
No
No
No
No
No
No
No
No
Manufacturer
Concentration Contact time
BODE Chemie 1.6%
Reprocessing
Steam
Special
sterilization
reprocessing
measures
No
No
No
No
No
No
No
No
Temperature
15 min
—
61