Repair - Dräger BiliLux Instructions For Use Manual

Led phototherapy light
Hide thumbs Also See for BiliLux:
Table of Contents

Advertisement

Service
12.6.3
Tightening the spring arm joints
1 Position the device so that the spring arm is in a horizontal position.
2 Remove the covers (A) on the spring arm joints.
3 Tighten the hex nut (B) in each joint using a 13 mm socket wrench so that the
device stays securely in the horizontal position. Do not overtighten.
4 Verify that the spring arm moves freely and remains in any position.
5 Reattach the covers on the spring arm joints.
A
12.7

Repair

Dräger recommends that all repairs are performed by DrägerService and that only
authentic Dräger repair parts are used.
68
B
A