6.4.2
Attaching the BiliLux to a GCX rail
1 Attach the pivot rail (A) to the GCX rail.
2 Attach the rail clamp for the phototherapy light (B) to the pivot rail.
3 Insert the post (C) on the bottom of the spring arm into the rail clamp (B). Ensure
that the arm is seated properly, and then tighten the adjustment screw (D) on the
rail clamp to secure the arm.
4 Insert the quick-connect plug (E) on the phototherapy light into the arm socket
(F) until it audibly clicks into place.
Instructions for use BiliLux
G
F
A
B
C
F
E
A
D
D
E
C
B
Assembly and preparation